The measuring chamber of the sensor is crossed by infrared radiation. When the air is introduced into this chamber, the molecules of interest absorb a part of the radiation, which reduces the intensity of the optical signal. The concentration of the gas can therefore be deduced according to the Beer-Lambert’s physical law.

NDIR N2O Sensor

Looking for custom gas measurement solutions for your project?

Technical specifications

OCIEngine PRO nitrous oxide datasheet
N2O DETECTION – MODEL NUMBER CO000033
Unit of measurement ppm
Measurement range 0 – 1000 ppm
Detection limit (3σ) 200 ppm
Repeatability 0.5% Full Scale + 1% reading
SAMPLING CHARACTERISTICS
Test duration 4000 ms minimum
Air inlet Minimal relative pressure at measuring cell inlet port: 0.3 millibars Maximal relative pressure at measuring cell inlet port: 10 millibars
Flow rate range From 0,1 to 1,5 L/min
Sampling period 125 ms
Sample temperature 34°C +5°C
Sample humidity >95% Human breath
ELECTRICAL CONDITIONS
Supply voltage 3,7 VDC min 5,5 VDC max
Input current 1,5 A max
Communication interface RS232 protocol (other on request) Solder pad connection (recommended cable sections : ≤0.13 mm² AWG ≥ 26)
Discover our evaluation kit to help you pre-integrate and test the sensor in your systems
ENVIRONMENTAL CONDITIONS
Operating conditions From 0°C to 50°C
Storage From -10°C to 70°C
Typical starting time 60s
PACKAGE INFORMATIONS
Dimensions and weight 21 x 21 x 80 mm, 17g
External plastic casing ABS PC-UL94V0
Fixing parts 3 apertures (Ø 1.6 mm) compatible with self-tapping screws
(1.8 mm recommended) to fix OCIEngine