OCISense is an advanced infrared spectrometer designed for the accurate analysis of Volatile Organic Compounds (VOC) in both exhaled and ambient air.  This OCISense is designed to detect and quantify CO2, ethanol and N2O molecules in breath air, offering precious insights into respiratory health and metabolic processes.

 

The measurement chamber of the sensor is ingeniously designed to be crossed by infrared radiation. As exhaled air flows through this chamber, the molecules of interest selectively absorb specific wavelengths of the radiation, resulting in a perceptible reduction in the optical signal. Based on the principles of Beer-Lambert’s law, the sensor accurately deduces the concentrations of these gases, providing essential data to help medical diagnosis.

OCISense Breath

CO2

Ethanol

N2O

For any inquiry, please

Technical specifications

CO2, EtOH & N2O DETECTION – MODEL NUMBER CO000036

GAS DETECTION
EtOH CO2 N2O
Unit of measurement µg/l % ppm
Measuring ranges 0 – 2000 µg/l 0 – 7% 0 – 1000 ppm
Accuracy 0 to 400 μg/l: 7 μg/l 400 to 2000 μg/l: ±1.6% measure 0.01% + 2% measure 0,5% FS + 2% measure
Resolution 1µg/l TBD TBD
Other ethanol specific features based on the OIML R126 standard:
Measurement error 0 to 400 μg/l: 20µg/l 400 μg/l to 2000 μg/l: ±5% measure
Drift < 10 μg/l for a reference of 0 μg/l in 4h < 10 μg/l for a reference of 400 μg/l in 4h < 20 μg/l for a reference of 400 μg/l in 2 months
Memory effect < 10 μg/l for alternative tests with a reference concentration > 1 500 μg/l and 100 μg/l
SAMPLING CHARACTERISTICS
Sampling period 125 ms
Response time 1s (τ=63%), at a flow rate of 1 L/min
Pressure range 300 – 1200 hPa (mbar)
Flow rate 0,1 to 1L/min
ELECTRICAL CHARACTERISTICS
Supply voltage 125 ms
Input current 1,5A max
Communication interface RS232
ENVIRONMENTAL CONDITIONS
Operating temperature0°C to 55°C
Storage temperature-25°C to +70°C
Starting time1min (30min at full specifications)
Relative humidity0-95% RH (non-condensing)
MECHANICAL CHARACTERISTICS 
Dimensions165 x 55 x 35 mm
WeightTBD
Fixing partsTBD

 